New SpettroREM research on hearing aid verification

Aurical IQ Freefit real ear measurement system for hearing aid fitting.

A peer-reviewed study published in Audiology Research in October 2026 tested the first step of SpettroREM, a method that analyzes how speech is shaped inside the ear canal during routine real ear measurement. Using recordings captured with the Aurical® Freefit, the researchers found that the broad spectral pattern of unaided speech in the ear canal could be measured with good repeatability across sessions. The complete method still needs aided vs. unaided studies before its clinical value can be established.

 

The study was led by Dr. Sandro Burdo and conducted by independent researchers in Italy. Audiologists at several clinical centers collected recordings using the same protocol and the Aurical Freefit probe microphone system with Otosuite® software.

Why look beyond output level

Real ear measurement (REM) uses a probe microphone in the ear canal to confirm that a hearing aid’s output matches the prescribed target. A 2021 systematic review in Trends in Hearing found that REM fittings were associated with better speech intelligibility and patient preference than manufacturer first-fit settings.

Modern hearing aids do more than amplify sound. Compression, directional microphones, noise reduction, and frequency lowering can all change how speech is organized over time and across frequencies. SpettroREM provides an objective, multidimensional measurement of speech characteristics in the ear canal, extending beyond output level and frequency to assess spectral and temporal properties of the speech signal.

 

How the SpettroREM method works

SpettroREM is designed as an objective measurement of the acoustic characteristics of speech in the ear canal. Calibration, probe tube placement, and stimulus presentation all stay with the REM system. SpettroREM records the speech signal from the probe microphone and runs it through an automated analysis in Praat, a free speech analysis program.

Each ear serves as its own baseline. The unaided recording from a patient’s ear canal is called the Individual Acoustic Reference (IAR). In the intended use, a clinician would compare an aided recording against that same ear’s IAR to see how the hearing aid changed the speech signal.

The analysis looks at 37 acoustic descriptors grouped into six areas:

  • Global spectral organization
  • Fricative characteristics, such as the /s/ sound
  • Vowel acoustics
  • Vowel geometry
  • Temporal envelope
  • Speech level and contrast

For the study, researchers played a 44-second Italian conversational dialogue with female, male, and child speakers from the Otosuite stimulus library through the REM loudspeaker. The signal from the Aurical Freefit analog monitoring output was captured with a USB audio interface and recorded in Praat. According to the authors, the unaided recording adds less than a minute of recording time, and the only extra hardware used was an audio interface.

 

What the study found

This study examined the test–retest reliability of the unaided IAR measurement, looking at how consistently the same measurement could be reproduced across separate recording sessions. Twenty-five participants had unaided recordings taken on two different days, with the probe tube removed and reinserted between sessions.

  • Global spectral measures were the most repeatable. The seven core descriptors had ICC values between 0.906 and 0.984.
  • Average differences between sessions in the three speech spectrum bands stayed under 1 dB.
  • A separate analysis of the participants’ left ears showed a similar pattern for the spectrum band levels.
  • Repeatability was less consistent for several vowel, fricative, temporal envelope, and level-dependent descriptors.

The researchers also tested the recording setup itself. Electrical testing showed that the cable, audio interface, and software did not meaningfully change the signal. The Otosuite display and the external recording did show frequency-dependent differences, so the authors advise treating them as separate measurements that are not interchangeable.

What the study does not show yet

The authors describe this as a preliminary step. The study did not compare aided and unaided recordings, so it does not yet show if SpettroREM can detect changes made by hearing aids or how those changes relate to speech understanding or patient benefit. The groupings of descriptors into reliability tiers were assigned after the analysis and need confirmation in a separate sample. Only one Italian speech stimulus has been tested so far. Next steps named by the authors include aided vs. unaided studies, testing with other languages and stimuli, and automated alignment of recordings.

Natus Sensory Academy

Read the full study

The complete paper, including the full descriptor results and technical validation, is available open access in Audiology Research. To learn more about the probe microphone system used in the study, explore Aurical Freefit.

Sources:

  1. Burdo, S., Carini, E., Impalà, G., & Brotto, D. (2026). Stability and Test–Retest Reliability of the Individual Acoustic Reference for Within-Subject Analysis of Hearing Device-Processed Speech: The SpettroREM Method. Audiology Research, 16(5), 158. https://doi.org/10.3390/audiolres16050158

054836 RevA