ICS Impulse vHIT – Performing LARP RALP Head Impulses
Overview
This tutorial demonstrates how to perform vertical semicircular canal testing during video head impulse testing (vHIT) using the LARP and RALP planes. It explains how to position and center the patient correctly, align the head to the appropriate canal plane, maintain fixation, and perform accurate vertical head impulses.
The session also covers head-position feedback, correct hand placement, impulse direction, velocity and amplitude, ROI positioning, and interpretation of software feedback. It highlights how proper alignment helps isolate the intended canals and reduce stimulation of the opposite test plane, supporting more accurate assessment of all six semicircular canals.
What you’ll learn
After watching this session, you will be able to:
- Position the patient correctly for LARP and RALP vHIT testing.
- Understand why the head is rotated approximately 35–45 degrees before performing vertical canal impulses.
- Maintain fixation on the target while minimizing unwanted torsional eye movement.
- Position the ROI correctly without repeating the original calibration.
- Use correct hand placement and perform small, high-velocity vertical head impulses.
- Recognize which impulse direction stimulates the anterior versus posterior canal.
- Use head-position feedback, training curves, and software messages to identify correctly and incorrectly performed impulses.
- Complete testing of the vertical canals as part of the full six-canal vHIT assessment.
Key takeaways
- LARP assesses the left anterior and right posterior canals, while RALP assesses the right anterior and left posterior canals.
- Rotating the head 35–45 degrees helps align the intended vertical canal pair with the test plane and improves impulse accuracy.
- The patient should continue looking at the fixation target even while the head is rotated.
- Vertical head impulses should remain small in amplitude, approximately 10–15 degrees, while maintaining sufficient velocity.
- Downward impulses primarily stimulate the anterior canal, while backward impulses stimulate the posterior canal.
- The software’s head-position feedback and training curves help confirm whether the correct plane is being stimulated.
- A “wrong plane stimulated” message indicates that the impulse was directed incorrectly or that the head drifted toward center.
- Once lateral, LARP, and RALP testing are complete, data have been collected from all six semicircular canals.